Bubble Paddle Data
 

     
 
S/2 S/4 S/8 S/16 S/32 S/64 S/128 S/256 S/512 S/1024 S/2048 S/4096
Sample 1 2.1720 1.8192 1.4273 1.0231 0.6934 0.3893 0.2045 0.1141 0.0696 0.0459 0.0345 0.0289
Sample 2 2.1729 1.8458 1.4432 1.0588 0.6964 0.3933 0.2131 0.1166 0.0708 0.0465 0.0343 0.0286
Sample 3 2.1806 1.8463 1.4563 1.0723 0.6844 0.3983 0.2164 0.1200 0.0739 0.0467 0.0352 0.0292
Sample 4 2.2276 1.8544 1.4625 1.0960 0.6981 0.3995 0.2207 0.1192 0.0716 0.0469 0.0357 0.0289
Sample 5 2.2446 1.9103 1.5122 1.1458 0.7445 0.4309 0.2338 0.1266 0.0773 0.0499 0.0371 0.0304
                       
Average 2.1995 1.8552 1.4603 1.0792 0.7034 0.4023 0.2177 0.1193 0.0726 0.0472 0.0354 0.0292
Stdev 0.034076 0.033542 0.031999 0.045631 0.023597 0.016521 0.010785 0.004694 0.003042 0.001566 0.001122 0.000704
%CV 1.55% 1.81% 2.19% 4.23% 3.35% 4.11% 4.95% 3.93% 4.19% 3.32% 3.17% 2.41%
 


     
 
Samples were removed from a V&P bead suspension bath at different locations.
The samples were dispensed into a reader plate, serially diluted, and read for absorbance.
     
 
Sample 1   Corner bottom (near paddle wheel spindles)
Sample 2   Against length wall, bottom of bath
Sample 3   Center of bath, top of liquid
Sample 4   Against length wall (near paddle wheel spindles - opposite side of Sample 2), bottom of bath
Sample 5   Corner bottom, adjacent to Sample 1
     
         